OXIDATION STABILITY TEST APPARATUS

Code: 11026-08


COMPLIANCE


Conforms to D5846; IP 48; IP 154; IP 280; IP 306; IP 307; IP 331 (obs); IP 335; EN 61125A, B and C; IEC 61125 Methods A, B and C; BS 2000 Parts 48, 280 and 306; ISO 2160; ISO 7624; DIN 51811; NF T60-150; JIS K2513 IS:335/1993; ASTM D130; D2440; D4636; D4871;D5763; with suitable change of flow meters of the appropriate flow rate and other small changes.


APPLICATION


Oxidation Stability test is one of the most basic tests to determine the shelf /working life of any petroleum product.

We are also available on Whatsapp between 11 am to 6 pm

INSTRUMENT AND CONSTRUCTION


Consists of a solid aluminum block bath to accommodate 4/6/12/24 test vessels. with Digital Micro Processor based Temp control with pt-100 sensor.Operation on 230 V, 50Hz., Single phase, AC Mains. Bath temp. Range 50°C to 375°C., Accuracy + 0.10C. Complete with required Nos. of Oxidation tube(s) made of ‘Borosil’ Glass.


SPECIFICATIONS



* 7" Colour Touch screen based command and control system with :-


* Automatic control of air/oxygen flow rate with digital display.


* Automatic “TAMPER PROOF” Data recording with recording of flow, temperature, time and faults during testing.


* Auto Diagnostic features like NO GAS FLOW, HEATER FAILURE, FLOW FAILURE CYLINDER LEVEL LOW, SENSOR FAILURE ETC.


* Password protection for system settings.


* Internet/Lan/PC connectivity for seamless operation and monitoring from any where in the world.


* Flow rates from 0.15 LPH to 3.2LPH


* Single system for both Air/Oxygen. User selectable


* 100% GLP compliance


ECO-FRIENDLY FEATURES


1. Auto cut-off of gas flow at the end of test timing.


2. Auto cutoff of heaters at the end of test timing.


3. Paper less recording with inbuilt 2GB memory, hence not required to be always connected to a PC/Printer


4. Special low density heaters for longer life and lesser power consumption.


5. Can be configured for multiple independent test blocks as per requirement, Max 24 tests.



STANDARD ACCESSORIES


11026-21 Test Vessel


11026-22 Copper Catalyst


11026-23 Analog Flow meter 1LPH O2 Silicone Tubing

OPTIONAL ACCESSORIES & SPARES


11026-21 Spare Test Vessel


11026-22 Spare Copper Catalyst


11026-23 Analog Flow meter for Oxygen 1LPH


11026-24 Mandrel for Catalyst preparation


11026-25 Test vessel holding stand with tubing Double Stage cylinder pressure regulator.


P.CODE         MODELS                 DESCRIPTION


11026-00           OSA-T4                        4 Test block - Basic


11026-01           OSA-T6                        6 Test block - Basic


11026-02           OSA-T8                        8 Test block - Basic


11026-03           OSA-T12                     12 Test block - Basic


11026-04           OSA-T24                     24 Test block - Basic


11026-05           OSA-T4A                      4 Test Automatic


11026-06           OSA-T6A                      6 Test Automatic


11026-07           OSA-T12A                    12 Test Automatic


11026-08           OSA-T24A                    24 Test Automatic